Audio-visual alarm facility for Overheating beyond set temperature range, Spurt in Heater output i.e., more than 60% heater output for 10 minutes system should raise alarm and cut-off heater, Power failure, Heater failure, Patient temperature less/Greater than required/set temperature, Probe failure
Observation light Luminance and Color temperature > = 1000 Lux, Color temperature range from 3700K to 5100K
Battery backup facility for Power failure indication during power fail Yes
Desired Temperature Range: 25 To 40 Degree C
Settable Temperature : 32 To 38 Deg C
Temperature Resolution : <= 0.1 Degree C
Temperature Accuracy: <= 0.2 Degree C
Overall Size: 600L x 450W x 1800H mm
Noise: Auditory alarm shall have a sound level of at least 65 dBA at a distance of 3 m from the front of the infant radiant warmer, and the sound level of the alarm shall not exceed 80 dBA on the mattress